ITC Field Guide
ITC/GUIDE
A working guide for PhD researchers · biophysics · structural biology · medicinal chemistry

Isothermal Titration Calorimetry

ITC measures the heat released or absorbed each time a small volume of ligand is injected into a cell holding its binding partner. Fit those heats to a binding model and one experiment returns the affinity, the enthalpy and the stoichiometry. The free energy and entropy follow from them, with no labels and no immobilisation.

Ka · Kdfitted · curve shape
ΔHfitted · curve amplitude
nfitted · inflection position
ΔGderived · −RT ln Ka
−TΔSderived · ΔG − ΔH
Live trace · 1 µM Kd · ΔH −10 kcal/mol · 19 injectionsDP (µcal/s)
§1 · Principle

Heat is the universal signal

Almost every binding event exchanges heat with its surroundings. ITC counts that heat directly, so the measurement needs no chromophore, no fluorophore and no surface.

The cell holds the macromolecule M (often a protein, at roughly 5–100 µM). A syringe injects the ligand L in 15–30 small aliquots. At first nearly every injected ligand molecule binds, so each injection releases (or absorbs) almost the full molar enthalpy. As free sites run out, less ligand binds per injection and the peaks shrink until only the heat of dilution remains.

The heat of injection i is proportional to how much new complex it forms:

Heat per injection\[ q_i = \Delta H \, V_0 \left( [ML]_i - [ML]_{i-1} \right) \tag{1} \]

The shape of the resulting isotherm encodes the association constant. Affinity and enthalpy then give the rest of the thermodynamic profile:

Free energy (standard state c° = 1 M)\[ \Delta G^\circ = -RT\ln\!\left(K_a c^\circ\right) = RT\ln\!\left(\frac{K_d}{c^\circ}\right) \tag{2} \]
Gibbs–Helmholtz partition\[ \Delta G^\circ = \Delta H^\circ - T\Delta S^\circ \tag{3} \]

Sign convention. Negative ΔH means heat is released (exothermic). Negative −TΔS means entropy favours binding. At 25 °C, RT = 0.593 kcal/mol, so every tenfold gain in affinity is worth 1.36 kcal/mol of ΔG.

What ΔH and ΔS tell you (with care)

TermTypical molecular contributors
ΔH < 0Hydrogen bonds, van der Waals contacts, salt bridges formed in the complex, minus the cost of desolvating polar groups
−TΔS < 0Release of ordered water from apolar surfaces (hydrophobic effect), counter-ion release
−TΔS > 0Loss of rotational, translational and conformational freedom; folding or ordering on binding

ΔH and ΔS are global sums over protein, ligand, water, ions and buffer. Treat structural interpretations of a single signature as hypotheses. Compare series of related ligands instead.

Where ITC fits among binding methods

ITCSPR / BLIFluorescence
Label or surfaceNoneImmobilised partnerProbe or intrinsic Trp
Gives ΔH, nDirectlyNo (van 't Hoff only)No
KineticsLimitedkon, koffStopped-flow
Sample useHigh (≈ 10–300 µg)LowLow
Direct Kd window≈ 10 nM – 100 µMpM – mMpM – mM
§2 · Instrument

Two cells, one feedback loop

A power-compensation calorimeter keeps the sample cell at the same temperature as a reference cell. The raw signal is the extra power that the feedback heater has to supply.

ADIABATIC JACKET · isothermal shield Reference cell buffer or water Sample cell · M Syringe · ligand L ≈ 10–20× cell conc. needle + stirring paddle (≈ 750 rpm) ΔT sensor held at ΔT = 0 constant reference power feedback power → DP Exothermic binding heats the sample cell, so the feedback heater supplies less power. DP drops: peaks point down. DP
Schematic of a power-compensation ITC. Cells are coin-shaped and completely filled; each injection pushes an equal volume out of the active cell volume V0 through the neck. Heat-conduction designs (e.g. TA Nano ITC) measure the same heat through a thermopile instead.

Differential power (DP) is the signal. It is reported in µcal/s or µW (1 µcal/s = 4.184 µW). Integrating each peak over time gives the heat of that injection in µcal.

Heat of injection i\[ q_i = \int_{t_i}^{t_{i+1}} \left[\,\mathrm{DP}(t) - \mathrm{DP}_{\text{baseline}}(t)\,\right] dt \tag{4} \]

The instrument's response time sets how long a peak takes to decay (τ ≈ 10–20 s). The spacing between injections must be long enough for the signal to return to baseline, which slow binding or conformational steps extend.

Typical run parameters

ParameterLow-volume cellStandard cell
Cell volume V0≈ 200 µL≈ 1.4 mL
Syringe volume≈ 40 µL≈ 280 µL
Injections1 × 0.4 + 18 × 2 µL1 × 2 + 27 × 10 µL
Spacing120–180 s180–300 s
Stirring750 rpm300–310 rpm
Smallest usable heat≈ 0.1–0.5 µcal≈ 0.5–1 µcal

Examples: MicroCal PEAQ-ITC and iTC200 (low volume), VP-ITC (standard). Always check your instrument's manual for exact volumes.

§3 · Interactive simulator

Run a titration before you run a titration

Change the affinity, enthalpy, stoichiometry and concentrations and watch the thermogram and the integrated isotherm respond. The simulator uses the exact single-site model with displaced-volume correction (equations 6–9 in §4).

Scenarios
Thermogram · raw DPexothermic peaks point down
Integrated isotherm · heat per mole of injectant○ first injection, discarded
Wiseman c-value · c = n Ka [M]
0.010.111010010³10⁴10⁵
Thermodynamic signature (kcal/mol)
§4 · The binding model

From mass balance to the Wiseman isotherm

Most ITC data are fitted to n identical, independent sites. Every step below is what fitting software does for you, so it pays to know where each parameter enters.

1 · Equilibrium and mass balance

Association constant and conservation of mass\[ K_a = \frac{1}{K_d} = \frac{[ML]}{[M]_{\text{free}}[L]_{\text{free}}},\qquad n[M]_t = [M]_{\text{free}} + [ML],\qquad [L]_t = [L]_{\text{free}} + [ML] \tag{5} \]

Here [M] counts binding sites through n. Substituting gives a quadratic in [ML] with one physical root:

Exact complex concentration\[ [ML] = \tfrac12\left( n[M]_t + [L]_t + K_d - \sqrt{\left(n[M]_t + [L]_t + K_d\right)^2 - 4n[M]_t[L]_t}\,\right) \tag{6} \]

2 · Total heat after injection i

With Xt for total ligand and Mt for total macromolecule in the active volume V0, the cumulative heat is (Wiseman et al., 1989):

Cumulative heat\[ Q_i = \frac{n M_t \Delta H V_0}{2}\left[1 + \frac{X_t}{nM_t} + \frac{1}{nK_aM_t} - \sqrt{\left(1 + \frac{X_t}{nM_t} + \frac{1}{nK_aM_t}\right)^{2} - \frac{4X_t}{nM_t}}\;\right] \tag{7} \]

3 · Correcting for the overfilled cell

Each injection δVi expels the same volume from the active cell. After a cumulative volume ΔV, concentrations are diluted:

Concentrations in the active volume\[ M_t = M_0\,\frac{1 - \frac{\Delta V}{2V_0}}{1 + \frac{\Delta V}{2V_0}}, \qquad X_t = X_0\,\frac{\Delta V}{V_0}\left(1 - \frac{\Delta V}{2V_0}\right) \tag{8} \]
Heat measured for injection i\[ \Delta Q_i = Q_i + \frac{\delta V_i}{V_0}\left(\frac{Q_i + Q_{i-1}}{2}\right) - Q_{i-1} \tag{9} \]

The isotherm plots ΔQi/(δViX0), the heat per mole of injectant, against the molar ratio Xt/Mt.

4 · The c-value decides what you can measure

\[ c = nK_a M_t = \frac{n M_t}{K_d} \tag{10} \]

For n = 1 and small injections, the isotherm has a universal shape set only by c and the molar ratio XR = Xt/Mt:

\[ \frac{1}{V_0}\frac{dQ}{dX_t} = \Delta H\left[\frac12 + \frac{1 - X_R - 1/c}{2\sqrt{\left(1 + X_R + 1/c\right)^2 - 4X_R}}\right] \tag{11} \]
  • c ≈ 10–100: a clear sigmoid. Ka, ΔH and n all come out well determined.
  • c > 1000: a step function. ΔH and n are exact, Ka only has a lower bound. Use a displacement titration.
  • c < 1: a shallow curve with no inflection. Ka is still measurable if the titration reaches high saturation (XR ≫ 1) and n is fixed from independent knowledge (Turnbull & Daranas, 2003).

5 · Displacement for very tight binders

Pre-saturate M with a weak competitor B of known KB and ΔHB, then titrate the tight ligand A. The apparent parameters fall back into the measurable window (Sigurskjold, 2000; Velazquez-Campoy & Freire, 2006):

\[ K_{\text{app}} = \frac{K_A}{1 + K_B[B]}, \qquad \Delta H_{\text{app}} = \Delta H_A - \Delta H_B\,\frac{K_B[B]}{1 + K_B[B]} \tag{12} \]

6 · Beyond one set of sites

  • Two independent sets: separate (n1, K1, ΔH1) and (n2, K2, ΔH2). Often biphasic isotherms.
  • Sequential sites: stepwise macroscopic constants K1, K2…; needed for cooperativity.
  • Dimer dissociation: inject concentrated protein into buffer to get Kdim and ΔHdim.
  • Reverse titration (M into L) tests the model independently and constrains n.

Add parameters only when residuals demand it and a replicate or orthogonal method supports the model.

§5 · Thermodynamics

Temperature reveals the heat capacity change

Repeat the titration at several temperatures. ΔH usually varies linearly with T, and its slope ΔCp reports on buried surface area and on water rearrangement.

Heat capacity change\[ \Delta C_p = \left(\frac{\partial \Delta H}{\partial T}\right)_p \tag{13} \]
Temperature dependence (constant ΔCp)\[ \Delta H(T) = \Delta H(T_0) + \Delta C_p\,(T - T_0), \qquad \Delta S(T) = \Delta S(T_0) + \Delta C_p \ln\frac{T}{T_0} \tag{14} \]
Empirical surface-area relation (Murphy & Freire, 1992)\[ \Delta C_p \approx 0.45\,\Delta A_{\text{apolar}} - 0.26\,\Delta A_{\text{polar}} \quad \left(\text{cal mol}^{-1}\,\text{K}^{-1},\ \Delta A \text{ in Å}^2\right) \tag{15} \]
van 't Hoff enthalpy\[ \frac{d\ln K_a}{d(1/T)} = -\frac{\Delta H_{\text{vH}}}{R} \tag{16} \]

Large negative ΔCp indicates burial of apolar surface. Because ΔH and TΔS both shift with temperature while ΔG barely moves, binding looks enthalpy-driven at one temperature and entropy-driven at another. This is the most common source of apparent enthalpy–entropy compensation. A mismatch between ΔHvH and the calorimetric ΔH points to coupled equilibria such as protonation or conformational change.

ΔG, ΔH and −TΔS versus temperature
§6 · Linked equilibria

Your buffer is part of the reaction

If binding takes up or releases protons, the buffer supplies or absorbs them, and its ionisation enthalpy adds to the heat you measure. Titrate in two or more buffers with different ionisation enthalpies to separate the terms.

Observed ΔH across buffers at the same pHslope = nH
Observed enthalpy\[ \Delta H_{\text{obs}} = \Delta H_{\text{bind}} + n_{H^+}\,\Delta H_{\text{ion}}^{\text{buffer}} \tag{17} \]

nH > 0 means the complex takes up protons from the buffer. Plot ΔHobs against ΔHion: the slope is nH and the intercept is the buffer-independent binding enthalpy.

BufferΔHion (kcal/mol, 25 °C)ΔHobs here

Ionisation enthalpies from Goldberg, Kishore & Lennen (2002), rounded. Phosphate and acetate hide proton linkage; Tris and imidazole magnify it. Tris also has a large temperature coefficient of pKa (≈ −0.028 per K).

§7 · Experimental design

Plan concentrations, then match everything

Most failed ITC runs fail in sample preparation. The planner below turns an expected Kd into cell and syringe concentrations using equations 8 and 10.

Samples

  • Dialyse or SEC the macromolecule into the final buffer. Dissolve or dilute the ligand in the same dialysate.
  • Match pH within 0.05 units and salt exactly. Match DMSO to within ≈ 0.1 % v/v; its heat of dilution is large.
  • Prefer TCEP (≤ 0.5 mM) to DTT or β-ME, which oxidise and cause baseline drift.
  • Centrifuge or filter to remove aggregates, and degas briefly to avoid bubbles.

Concentrations

  • Measure concentrations after dialysis (A280 with a calculated ε, or amino-acid analysis). Errors map directly into n and ΔH.
  • Syringe concentration is typically 10–20 × the cell concentration for n = 1, enough to reach a molar ratio of 2 or more.
  • Check the heat of the first real injections: aim for at least ≈ 5 µcal (low-volume) or 10–20 µcal (standard).

Controls and runs

  • Titrate ligand into buffer to measure the heat of dilution. Subtract it, or fit a constant offset.
  • Start with a small first injection (0.4 µL or 2 µL) and discard it: ligand diffuses from the needle tip during equilibration.
  • Wait for a flat baseline before the first injection. Run at least three independent replicates with fresh sample preparations.
§8 · Troubleshooting

Read the thermogram before you fit it

Most artefacts are visible in the raw DP trace. Try reproducing several of them in the simulator with the scenario buttons and the heat-of-dilution slider.

What you see
Likely cause
What to do
Large peaks that never shrink, identical from first to last injection
Buffer mismatch (pH, salt, DMSO, glycerol) or no binding at all
Re-dialyse, prepare ligand in dialysate, run ligand → buffer control
Step-like isotherm with only 1–2 points in the transition
c > 1000: binding too tight for the concentrations used
Lower [M] if the signal allows it, or switch to a displacement titration
Shallow, nearly linear decline with no plateau
c ≪ 1, or the titration stops far short of saturation
Raise concentrations or the syringe-to-cell ratio; fix n if it is known independently
n far from the expected value (e.g. 0.6 for a 1:1 complex)
Inactive or misfolded fraction, wrong ε or concentration, aggregation, or a true stoichiometry
Re-measure concentrations, check oligomeric state by SEC-MALS or DLS, run the reverse titration
Spikes and a noisy baseline
Air bubbles, dirty cell, bent syringe or paddle
Degas, clean the cell with detergent, reload carefully, inspect the syringe
Drifting baseline
Incomplete equilibration, oxidising reducing agent, slow aggregation
Equilibrate longer, use TCEP, spin samples, check stability at the run temperature
Peaks fail to return to baseline before the next injection
Slow binding or a slow conformational change
Increase injection spacing; consider kinetic ITC analysis
Heats change sign during the titration
Two processes with opposite ΔH (two site classes, dimer dissociation, ligand self-association)
Test a two-site or coupled model, dilute each component alone, run the reverse titration
§9 · Analysis and reporting

From peaks to publishable numbers

Workflow

  1. Integrate peaks. Use a consistent baseline; automated shape-based integration (NITPIC) reduces bias and estimates per-injection errors.
  2. Handle dilution heat. Subtract the ligand → buffer control, or fit a constant offset if the late-injection heats are flat.
  3. Choose the simplest model consistent with independent knowledge of stoichiometry and oligomeric state.
  4. Fit by nonlinear least squares (Levenberg–Marquardt), excluding the first injection.
  5. Inspect residuals. Systematic trends mean the wrong model, a wrong concentration or a baseline problem.
  6. Replicate and fit globally. Fit replicates, reverse titrations or several temperatures together (SEDPHAT, AFFINImeter, PyTC).
  7. Report the uncertainty across independent replicates, not only the fit's standard error.

How concentration errors propagate

If this is wrong by factor fnΔHKd
Cell [M]× funchanged≈ unchanged
Syringe [L]× 1/f× faffected

Here f = (true concentration) / (assumed concentration). A fitted n of 0.7 therefore flags the cell concentration, or the active fraction, before it flags the biology.

Minimum reporting checklist

  • Instrument, cell volume, temperature, stirring speed, injection scheme
  • Buffer, pH, salt, additives, DMSO; how samples were matched
  • Cell and syringe concentrations and how they were measured
  • c-value, final molar ratio, model used, which parameters were fixed
  • n, Kd, ΔH, ΔG, −TΔS as mean ± SD of N independent runs
  • A representative thermogram and isotherm with fit and residuals
§10 · Self-check

Six questions before you book the instrument

0 of 6 answered

§11 · Reference

Symbols and further reading

SymbolMeaningTypical units
Ka, KdAssociation and dissociation constants, Kd = 1/KaM⁻¹, M
ΔHBinding enthalpy per mole of ligand boundkcal/mol, kJ/mol
ΔG, ΔSGibbs energy and entropy of bindingkcal/mol, cal/mol·K
nApparent stoichiometry, sites per molecule in the cell—
cWiseman parameter, n Ka Mt—
V0Active cell volumeµL, mL
M0, X0Initial cell and syringe concentrationsµM
Mt, XtTotal concentrations in V0 after injection iµM
δVi, ΔVInjection volume and cumulative injected volumeµL
DPDifferential power (raw signal)µcal/s, µW
ΔCpHeat capacity change of bindingcal/mol·K
nHProtons taken up per binding event—
RGas constant1.987 cal/mol·K
  1. Wiseman T, Williston S, Brandts JF, Lin LN (1989). Rapid measurement of binding constants and heats of binding using a new titration calorimeter. Anal. Biochem. 179, 131–137.
  2. Turnbull WB, Daranas AH (2003). On the value of c: can low affinity systems be studied by isothermal titration calorimetry? J. Am. Chem. Soc. 125, 14859–14866.
  3. Sigurskjold BW (2000). Exact analysis of competition ligand binding by displacement isothermal titration calorimetry. Anal. Biochem. 277, 260–266.
  4. Velazquez-Campoy A, Freire E (2006). Isothermal titration calorimetry to determine association constants for high-affinity ligands. Nat. Protoc. 1, 186–191.
  5. Goldberg RN, Kishore N, Lennen RM (2002). Thermodynamic quantities for the ionization reactions of buffers. J. Phys. Chem. Ref. Data 31, 231–370.
  6. Murphy KP, Freire E (1992). Thermodynamics of structural stability and cooperative folding behavior in proteins. Adv. Protein Chem. 43, 313–361.
  7. Freyer MW, Lewis EA (2008). Isothermal titration calorimetry: experimental design, data analysis, and probing macromolecule/ligand binding and kinetic interactions. Methods Cell Biol. 84, 79–113.
  8. Keller S, Vargas C, Zhao H, Piszczek G, Brautigam CA, Schuck P (2012). High-precision isothermal titration calorimetry with automated peak-shape analysis. Anal. Chem. 84, 5066–5073.
  9. Chaires JB (2008). Calorimetry and thermodynamics in drug design. Annu. Rev. Biophys. 37, 135–151.